作者:李巧元1项美香2宿燕岗3戴文龙1何东方1方冬平1郭成军1
英文作者:
单位:1首都医科大学附属北京安贞医院心内六科100029;2浙江大学医学院附属第二医院心内科,杭州310009;3复旦大学附属中山医院心内科,上海200032
英文单位:
关键词:磁共振成像;心脏再同步化治疗及埋藏式心脏自动除颤器;兼容磁共振成像心脏起搏器
英文关键词:
【摘要】目的 观察植入兼容磁共振成像(MRI)心脏再同步化治疗及埋藏式心脏自动除颤器(CRT-D)围术期以及术后MRI检查的安全性。方法 2017年12月至2018年3月选取药物控制欠佳的心力衰竭患者12例,植入兼容MRI的CRT-D(Inventra 7 HF-T QP),观察植入时、植入后1个月、MRI检查前后、MRI检查后1个月及3个月起搏器参数等情况。结果 12例植入CRT-D患者中男7例、女5例,10例(83.3%)患者进行了MRI检查,CRT-D植入时、植入后1个月、MRI检查前、MRI检查后、MRI检查后1个月及3个月电学参数差异均无统计学意义(均P>0.05)。MRI检查后1个月与MRI检查前相比,右心房阻抗的变化为(51±31)Ω,左心室阻抗的变化为(116±90)Ω,右心室阻抗的变化为(26±4)Ω,除颤阻抗的变化为(5.6±2.1)Ω。其中3例存在伪影,但不影响阅片诊断,1例心脏区伪影严重影响阅片;所有患者均未见起搏阈值增高、感知异常及导线移位、心肌穿孔等情况。结论 兼容MRI的CRT-D植入是安全可行的,植入后能安全进行MRI扫描。
【Abstract】Objective To observe the safety of magnetic resonance imaging(MRI)-conditional cardiac resynchronization therapy, implantable cardiac defibrillator(CRT-D) implantation and postoperative MRI performance. Methods From December 2017 to March 2018, 12 heart failure patients with poor response to medication therapy were implanted MRI-conditional CRT-D(Inventra 7 HF-T QP). Pacemaker parameters were observed during implantation, 1 month after operation, before MRI scanning and 1, 3 months after MRI examination. Results There were 7 males and 5 females. MRI was performed in 10 cases(83.3%). There were no significant changes of CRT-D electrical parameters during implantation, 1 month after operation, before and 1, 3 months after MRI scanning(all P>0.05). The variations of electrical parameters at 1 month after MRI were (51±31)Ω for right atrial impedance, (116±90)Ω for left ventricular impedance, (26±4)Ω for right ventricular impedance and (5.6±2.1)Ω for defibrillation impedance. Three cases showed slight artifacts in MRI which did not affect diagnosis, while 1 case had severe artifact around the heart that hindered image reading. No threshold increasing, sensation abnormality and lead dislodgment, myocardium perforation were observed. Conclusion MRI-conditional CRT-D pacemaker is feasible and MRI scanning can be performed safely after implantation.
copyright
地址:北京市朝阳区安贞路2号首都医科大学附属北京安贞医院北楼二层
电话:010-64456116 传真:010-64428528 邮编:100029 Email: zgyy8888@163.com
网址: 京ICP备2020043099号-3
当您在使用本网站投稿遇到困难时,请直接将稿件投送到编辑部邮箱zgyy8888@163.com。